کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4441008 1311087 2009 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Inter-comparison of three instruments for measuring regional background carbon monoxide
موضوعات مرتبط
مهندسی و علوم پایه علوم زمین و سیارات علم هواشناسی
پیش نمایش صفحه اول مقاله
Inter-comparison of three instruments for measuring regional background carbon monoxide
چکیده انگلیسی

Three CO instruments, i.e., vacuum ultraviolet resonance fluorescence (VUV-RF), gas chromatography/reduction gas detection (GC/RGD), and non-dispersive infrared (NDIR) spectrometry, were inter-compared at a background site, i.e., Lulin Atmospheric Baseline Station (LABS) in central Taiwan. For a period of 18 days of continuous measurements, highly coherent results were observed between the three instruments. Calibration of the GC/RGD and VUV-RF was based on the same batch of CO standards, producing results that agreed within 4%. Nevertheless, NDIR displayed a significant bias with a large offset compared to the other two instruments (about −63.1 ppb), due mainly to its over-simplified calibration procedure. The NDIR bias was then removed by aligning the NDIR data to those of GC/RGD in a process termed “post-run correction”. Through this practice, the NDIR was linked to the same calibration scale as that of GC/RGD. As a result, by simultaneously exploiting both GC/RGD and NDIR to measure background CO, both time-resolution and accuracy can be secured.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Atmospheric Environment - Volume 43, Issue 40, December 2009, Pages 6449–6453
نویسندگان
, , , , , , ,