کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
5407526 1393220 2007 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Object dependent sweep width reduction with spectral-spatial EPR imaging
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی تئوریک و عملی
پیش نمایش صفحه اول مقاله
Object dependent sweep width reduction with spectral-spatial EPR imaging
چکیده انگلیسی
For spectral-spatial EPR imaging, prior knowledge about the spatial support of an imaged object can be exploited in two ways. We can shrink the spatial field of view (FOV) to closely wrap the object in a sphere or reduce the sweep width in a projection dependent fashion. Use of a smaller spatial FOV with the same number of samples enhances spatial resolution by reducing voxel volume at the expense of signal-to-noise and a consequent degraded line-width resolution. We have developed another approach to define sweep width that prunes away the portions of the projection sweep with no signal. This reduces data acquisition time for the continuous wave (CW) EPR image proportional to the sweep width reduction. This method also avoids voxel volume reduction. Using the reduced-sweep method, we decreased the data acquisition time by 20% maintaining spatial and linewidth resolution.