We have developed a miniature double-pass cylindrical mirror electron energy analyzer (DPCMA) with an outer diameter of 26Â mm. The DPCMA consists of a shield for the electric field, inner and outer cylinders, two pinholes with a diameter of 2.0Â mm, and an electron multiplier. By assembling the DPCMA in a coaxially symmetric mirror electron energy analyzer (ASMA) coaxially and confocally we developed an analyzer for Auger photoelectron coincidence spectroscopy (APECS). The performance was estimated by measuring the Si-LVV-Auger Si-1s-photoelectron coincidence spectra of clean Si(1Â 1Â 1). The electron-energy resolution of the DPCMA was estimated to be E/ÎEÂ =Â 20. This value is better than that of the miniature single-pass CMA (E/ÎEÂ =Â 12) that was used in the previous APECS analyzer.
Journal: Surface Science - Volume 601, Issue 17, 1 September 2007, Pages 3589-3592