کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
5434919 1509148 2017 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A numerical study on the application of the functionally graded materials in the stent design
ترجمه فارسی عنوان
یک مطالعه عددی در مورد استفاده از مواد طبقه بندی عملکردی در طراحی استنت
موضوعات مرتبط
مهندسی و علوم پایه مهندسی مواد بیومتریال
چکیده انگلیسی


- FGMs showed a lower dogboning compared to the stents made of the uniform materials.
- The stent with the lowest heterogeneous index displayed the lowest dogboning.
- Steel stent showed the lowest foreshortening and fully expansion pressure.
- FGM with the heterogeneous index of 0.5 expected to exhibit the most suitable results.
- The results suggest the importance of the material parameters on the stent deformation.

Undesirable deformation of the stent can induce a significant amount of injure not only to the blood vessel but also to the plaque. The objective of this study was to reduce/minimize these undesirable deformations by the application of Functionally Graded Materials (FGM). To do this, Finite Element (FE) method was employed to simulate the expansion of a stent and the corresponding displacement of the stenosis plaque. Three hyperelastic plaque types as well as five elastoplastic stents were simulated. Dogboning, foreshortening, maximum stress in the plaque, and the pressure which is needed to fully expand the stent for different stent materials, were acquired. While all FGMs had lower dogboning in comparison to the stents made of the uniform materials, the stent with the lowest heterogeneous index displayed the lowest amount of dogboning. Steel stent showed the lowest foreshortening and fully expansion pressure but the difference was much lower than that the one for dogboning. Therefore, the FGM with the heterogeneous index of 0.5 is expected to exhibit the most suitable results. In addition, the results revealed that the material parameters has crucial effects on the deformation of the stent and, as a result, as a design point of view the FGM parameters can be tailored to achieve the goal of the biomechanical optimization.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Materials Science and Engineering: C - Volume 73, 1 April 2017, Pages 182-188
نویسندگان
, , , ,