کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
5497842 | 1532071 | 2017 | 4 صفحه PDF | دانلود رایگان |
عنوان انگلیسی مقاله ISI
Development of a windowless multi-wire proportional chamber (MWPC) counting system for measuring extended-area beta source
دانلود مقاله + سفارش ترجمه
دانلود مقاله ISI انگلیسی
رایگان برای ایرانیان
کلمات کلیدی
موضوعات مرتبط
مهندسی و علوم پایه
فیزیک و نجوم
تشعشع
پیش نمایش صفحه اول مقاله
چکیده انگلیسی
A windowless MWPC counting system has been developed to measure surface emission rate of extended-area sources. A new anode and guard plane design gives better electric-field uniformity in the (250Ã200)Â mm2 sensitive volume relative to a previous-generation MWPC at KRISS, which has been used as a national standard with uncertainty <1%. The conditions of stable operation for the new MWPC system were studied, and the uncertainty was estimated with an extended-area 36Cl source. The relative combined standard uncertainty was evaluated to be 0.23%, a 50% improvement over the previous-generation MWPC counting system.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 126, August 2017, Pages 175-178
Journal: Applied Radiation and Isotopes - Volume 126, August 2017, Pages 175-178
نویسندگان
S.H. Hwang, J.M. Lee, K.B. Lee, T.S. Park,