Improvement of the WBC calibration of the Internal Dosimetry Laboratory of the CDTN/CNEN using the physical phantom BOMAB and MCNPX code
چکیده انگلیسی


- Calibration process using the BOMAB physical phantom.
- Calibration of the efficiency curves of the Whole Body Counter in operation at the LDI/CDTN.
- Monte Carlo simulations BOMAB phantom.

The Laboratory of Internal Dosimetry of the Center for Development of Nuclear Technology (LDI/CDTN) is responsible for routine internal monitoring of occupationally exposed individuals. The determination of photon emitting radionuclides in the human body requires calibration of the detector in specific counting geometries. The calibration process uses physical phantoms containing certified activities of the radionuclides of interest. The objective of this work was to obtain calibration efficiency curves of the Whole Body Counter in operation at the LDI/CDTN using a BOMAB physical phantom and Monte Carlo simulations.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 117, November 2016, Pages 123-127