The quenching of sensitivity of gafchromic films EBT3 with increasing radiation LET is studied and a proposed method for correcting the measured dose images is described. The method evaluates a correction coefficient in each position of the acquired image by multiplying the measured dose value for a calculated ratio of the absorbed dose to the measured one. Such calculated doses are obtained utilising knowledge about absorbed and measured doses for single Bragg peaks from protons of a generic energy E. This topic is under study and the software in MATLAB developed for achieving the correction of the measured images is continuously implemented. The first results to test the validity of the proposed method also for complex dose distributions are promising.