کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
5499269 1533487 2017 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Commercial glass beads as TLDs in radiotherapy produced by different manufacturers
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Commercial glass beads as TLDs in radiotherapy produced by different manufacturers
چکیده انگلیسی
All bead types showed TL to be linear with dose (R2>0.999). Glow curve dosimetric peaks reached a maximum value at 300 °C for the Toho and 290 °C for the Czech and Mill Hill products but was between 200-250 °C for the Rocaille product. Radiation sensitivity following mass normalisation varied within an order of magnitude; Toho samples showed the greatest and Rocaille the least sensitivity. For the Toho, Czech, Rocaille and Mill Hill samples the energy responses at 80 kVp were 5.0, 4.0, 3.6 and 3.3 times that obtained at 6 MV. All four glass bead types offer potential use as TL dosimeters over doses commonly applied in radiotherapy. Energy response variation was <1% at 6 MV but significant variation was found for photon beam energies covering the kV range; careful characterisation is required if use at this range is intended.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiation Physics and Chemistry - Volume 137, August 2017, Pages 181-186