Very low-dose coronary artery calcium scanning with high-pitch spiral acquisition mode: Comparison between 120-kV and 100-kV tube voltage protocols
ترجمه فارسی عنوان
اسکن کلسیم عروق کرونری با دوز کم با حالت خرید مارپیچی بالا: مقایسه پروتکل های ولتاژ لوله 120 کیلو ولت و 100 کیلوولت
کلمات کلیدی
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
چکیده انگلیسی

BackgroundEffective radiation dose from a single coronary artery calcification CT scan can range from 0.8 to 10.5 mSv, depending on the protocol. Reducing the effective radiation dose to reasonable levels without affecting diagnostic image quality can result in substantial dose reduction in CT.ObjectivesWe prospectively compared tube voltages of 120 and 100 kV in a low-dose CT acquisition protocol for measuring coronary artery calcified plaque with prospectively electrocardiogram (ECG)-triggered high-pitch spiral acquisition.MethodsIn 150 consecutive patients, measurement of coronary artery calcified plaque was performed with prospectively ECG-triggered high-pitch spiral acquisition. Imaging was first done with tube voltage of 120 kV voltage and subsequently repeated with 100 kV and otherwise unchanged parameters. CT was performed with a dual-source CT system with 280 milliseconds of rotation time, 2 × 128 slices, pitch of 3.4, triggered at 60% of the R-R interval. Tube current for both protocols was set at 80 mAs. With the use of a medium sharp reconstruction kernel (Siemens B35f), cross-sectional images were reconstructed with 3.0-mm slice thickness and 1.5-mm increment. Agatston scores were determined per patient for both scan settings by 2 independent readers with the use of a standard threshold of 130 HU for calcium detection. In addition, the Agatston score was calculated with a previously proposed threshold of 147 HU for 100-kV acquisitions.ResultsMean image noise was 20 ± 5 and 27 ± 7 for 120 and 100 kV, respectively (P < 0.0001). Mean dose length product was 24 ± 6 cm · cGy for the 120-kV protocol and 14 ± 4 cm · cGy for the 100-kV protocol, corresponding to average estimated effective doses of 0.3 and 0.2 mSv (P < 0.0001). Five patients were excluded from the analysis. In the remaining 145 patients, using the standard tube voltage of 120 kV, any coronary calcium was detected in 76 identical patients by both observers. In 75 of these patients, calcium was also identified by both observers in 100-kV data sets, whereas 1 patient was scored negative by 1 reader and was assigned an Agatston score of 0.7 (threshold, 130 HU) and 0.2 (threshold, 147 HU) by the other. Interobserver disagreement for assigning a patient a zero Agatston score was the same for both scan settings (each 4 patients). The mean Agatston scores for 120-kV and 100-kV (threshold, 147 HU) scans were 105 ± 245 (range, 0-1865) and 116 ± 261 (range, 0-1917), respectively (P < 0.0001). Bland-Altman analysis indicated a systematic overestimation of the Agatston score with tube voltage of 100 kV and threshold of 147 HU (mean difference, 11; 95% limits of agreement, 62 to -40). Similar results were observed for coronary calcium volume scores.ConclusionHigh-pitch spiral acquisition allows coronary calcium scoring with effective doses below 0.5 mSv. The use of 100-kV tube voltage further reduces effective radiation dose compared with the standard of 120 kV; however, it leads to significant overestimation of the Agatston score when the standard threshold of 130 HU is used. Adjusting the threshold to 147 HU leads to a better agreement compared with standard 120 kV protocols yet with a remaining systematic bias toward overestimation of the Agatston score. For high-pitch spiral acquisition mode, effective radiation dose reduction when using a 100-kV setting is minimal compared with the standard 120-kV setting and may be considered nonsignificant in a clinical setting.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Cardiovascular Computed Tomography - Volume 7, Issue 1, January–February 2013, Pages 32-38
نویسندگان
, , , , , , , , , ,