کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
6040361 1188846 2006 11 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
EEG/(f)MRI measurements at 7 Tesla using a new EEG cap (“InkCap”)
موضوعات مرتبط
علوم زیستی و بیوفناوری علم عصب شناسی علوم اعصاب شناختی
پیش نمایش صفحه اول مقاله
EEG/(f)MRI measurements at 7 Tesla using a new EEG cap (“InkCap”)
چکیده انگلیسی
We aimed at improving the signal-to-noise ratio (SNR) of electroencephalography (EEG) during magnetic resonance imaging (MRI) by introducing a new EEG cap (“InkCap”) based on conductive ink technology. The InkCap was tested with temperature measurements on an electrically conductive phantom head and during structural and functional MRI (fMRI) recordings in 11 healthy human volunteers at 7 T. Combined EEG/fMRI measurements were conducted to study the interaction between the two modalities. The EEG recordings with the InkCap demonstrated up to a five-fold average decrease in signal variance during echo-planar imaging, with respect to a cap made of standard carbon fiber leads. During concurrent EEG/fMRI measurements in human volunteers, alpha oscillations were clearly detected at 7 T. Minimal artifacts were present in the T2* and high-resolution structural MR images of the brain parenchyma. Our results show that the InkCap technology considerably improves the quality of both EEG and (f)MRI during concurrent measurements even at 7 T.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: NeuroImage - Volume 33, Issue 4, December 2006, Pages 1082-1092
نویسندگان
, , , , , ,