کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
6052840 1196798 2016 43 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Accuracy of open-source software segmentation and paper-based printed three-dimensional models
ترجمه فارسی عنوان
دقت تقسیم بندی نرم افزار منبع باز و مدل های سه بعدی چاپ شده بر روی کاغذ
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی دندانپزشکی، جراحی دهان و پزشکی
چکیده انگلیسی
Using the optical scanner and specialized inspection software for measurements of accuracy of the created parts is recommended, as it allows one not only to measure 2-dimensional distances between anatomical points but also to perform more clinically suitable comparisons of whole surfaces. However, it requires specialized software and a very accurate scanner in order to be useful. Threshold-based, manually corrected segmentation of orbital walls performed with 3D Slicer software is accurate enough to be used for creating a virtual model of the orbit. The accuracy of the paper-based Mcor Matrix 300 3D printer is comparable to those of other commonly used 3-dimensional printers and allows one to create precise anatomical models for clinical use. The method of dividing the model into smaller parts and sticking them together seems to be quite accurate, although we recommend it only for creating small, solid models with as few parts as possible to minimize shift associated with gluing.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Cranio-Maxillofacial Surgery - Volume 44, Issue 2, February 2016, Pages 202-209
نویسندگان
, , ,