کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
6146969 1594958 2011 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Insulation failure in robotic and laparoscopic instrumentation: a prospective evaluation
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی پزشکی و دندانپزشکی (عمومی)
پیش نمایش صفحه اول مقاله
Insulation failure in robotic and laparoscopic instrumentation: a prospective evaluation
چکیده انگلیسی

ObjectiveThe purpose of this study was to detect the incidence, prevalence, and location of insulation failures (IFs) in laparoscopic and robotic instruments.Study DesignIn phase A, a total of 78 robotic and 298 laparoscopic instruments were tested at 20 W and 2.64 kV at Mayo Clinic in Arizona. In phase B, 60 robotic and 308 laparoscopic instruments were tested at 20 W/1 kV and 20 W/4.2 kV, respectively.ResultsIn phase A, the robotic group showed a higher prevalence (25/78; 32%) and incidence of IFs after 10 uses (35/44 instruments; 80%) when compared with laparoscopy (prevalence, 39/298 [13%]; incidence, 68/189 [36%]; P < .05). In phase B, IFs were detected in 81.7% of the robotic instruments and in 19.5% of the laparoscopic instruments (P < .005).ConclusionThere is a high incidence and prevalence of IF in endoscopic instrumentation that is more common in the robotic group.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: American Journal of Obstetrics and Gynecology - Volume 205, Issue 2, August 2011, Pages 121.e1-121.e5
نویسندگان
, , , ,