Validation of rebound tonometry for intraocular pressure measurement in the rabbit
ترجمه فارسی عنوان
اعتبار سنجی تنزل بازگشتی برای اندازه گیری فشار داخل چشم در خرگوش
موضوعات مرتبط
علوم زیستی و بیوفناوری ایمنی شناسی و میکروب شناسی ایمونولوژی و میکروب شناسی (عمومی)
چکیده انگلیسی


- TonoVet had a tendency to underestimate rabbit IOP compared to manometry.
- d mode is more accurate and precise than p mode.
- IOP recovered four days after cannulation in rabbits.
- Younger rabbits had lower IOP compared with older rabbits.
- Conversion tables for more accurate measurements have been presented.

Rabbits play a growing role in research into glaucoma surgical models and ocular drug delivery models. However, the lack of an accurate method for measuring intraocular pressure (IOP) in this animal has been a significant deficit. In this study we validated the use of the TonoVet rebound tonometer and provide conversion tables for its use in rabbits. Experiments were performed on 18 adult New Zealand White rabbits. The TonoVet measurements were obtained and compared to manometric readings by anterior chamber (AC) cannulation. The TonoVet position and 'd' (dog or cat) and 'p' (other species) modes were compared. The sensitivity of the TonoVet tonometer in assessing IOP changes was also tested. There was a strong linear correlation for both the 'd' mode (mean slope = 0.84 ± 0.03, r(2) = 0.99 ± 0.03) and the 'p' mode (mean slope = 0.64 ± 0.02, r(2) = 0.97 ± 0.01) of the TonoVet with manometric IOP. However, the TonoVet had a tendency to underestimate IOP compared to manometry and conversion formulae were possible to calculate for both modes. The orientation of the TonoVet handle had no effect on IOP reading, as long as the groove was horizontal. No significant differences were observed when comparing right and left eyes (P > 0.05). IOP recovered four days after cannulation. Younger rabbits had lower IOP compared with older rabbits (P < 0.01). Timolol produced a 2.5 mmHg reduction in IOP 2 h later as detected by the TonoVet. Using the conversion table presented, the TonoVet is a reliable and precise tool for the measurement of IOP in rabbits.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Experimental Eye Research - Volume 121, April 2014, Pages 86-93
نویسندگان
, , , , , ,