- We examine braking effect on COM progression exerted by an amputee's prosthetic foot.
- Use of Echelon hydraulic ankle-foot device was associated with reduced braking effect.
- As a result freely chosen walking speed was higher when using the device.
- Attenuation of a foot's braking effect may be as important as the energy it returns.
BackgroundIf a prosthetic foot creates resistance to forwards shank rotation as it deforms during loading, it will exert a braking effect on centre of mass progression. The present study determines whether the centre of mass braking effect exerted by an amputee's habitual rigid 'ankle' foot was reduced when they switched to using an 'Echelon' hydraulic ankle-foot device.MethodsNineteen lower limb amputees (eight trans-femoral, eleven trans-tibial) walked overground using their habitual dynamic-response foot with rigid 'ankle' or 'Echelon' hydraulic ankle-foot device. Analysis determined changes in how the centre of mass was transferred onto and above the prosthetic-foot, freely chosen walking speed, and spatio-temporal parameters of gait.FindingsWhen using the hydraulic device both groups had a smoother/more rapid progression of the centre of pressure beneath the prosthetic hindfoot (p â¤Â 0.001), and a smaller reduction in centre of mass velocity during prosthetic-stance (p < 0.001). As a result freely chosen walking speed was higher in both groups when using the device (p â¤Â 0.005). In both groups stance and swing times and cadence were unaffected by foot condition whereas step length tended (p < 0.07) to increase bilaterally when using the hydraulic device. Effect size differences between foot types were comparable across groups.InterpretationUse of a hydraulic ankle-foot device reduced the foot's braking effect for both amputee groups. Findings suggest that attenuation of the braking effect from the foot in early stance may be more important to prosthetic-foot function than its ability to return energy in late stance.
Journal: Clinical Biomechanics - Volume 29, Issue 7, August 2014, Pages 728-734