Agreement between temporal and spatial gait parameters from an instrumented walkway and treadmill system at matched walking speed
ترجمه فارسی عنوان
توافق بین پارامترهای فرامینی زمانی و مکانی از یک راهرو و سیستم تردمیل در سرعت پیاده روی
کلمات کلیدی
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
چکیده انگلیسی

BackgroundCommercially available instrumented treadmill systems that provide continuous measures of temporospatial gait parameters have recently become available for clinical gait analysis. This study evaluated the level of agreement between temporospatial gait parameters derived from a new instrumented treadmill, which incorporated a capacitance-based pressure array, with those measured by a conventional instrumented walkway (criterion standard).MethodsTemporospatial gait parameters were estimated from 39 healthy adults while walking over an instrumented walkway (GAITRite®) and instrumented treadmill system (Zebris) at matched speed. Differences in temporospatial parameters derived from the two systems were evaluated using repeated measures ANOVA models. Pearson-product-moment correlations were used to investigate relationships between variables measured by each system. Agreement was assessed by calculating the bias and 95% limits of agreement.ResultsAll temporospatial parameters measured via the instrumented walkway were significantly different from those obtained from the instrumented treadmill (P < .01). Temporospatial parameters derived from the two systems were highly correlated (r, 0.79-0.95). The 95% limits of agreement for temporal parameters were typically less than ±2% of gait cycle duration. However, 95% limits of agreement for spatial measures were as much as ±5 cm.ConclusionsDifferences in temporospatial parameters between systems were small but statistically significant and of similar magnitude to changes reported between shod and unshod gait in healthy young adults. Temporospatial parameters derived from an instrumented treadmill, therefore, are not representative of those obtained from an instrumented walkway and should not be interpreted with reference to literature on overground walking.

► We examined temporospatial parameters from two commercially available gait systems. ► Subjects walked at matched speed on a new instrumented treadmill and GAITRite mat. ► Limits of agreement were less than ±2% and ±5 cm for temporal and spatial parameters. ► Between-system differences were small but similar to those for shod and unshod gait. ► Gait parameters from the new treadmill are not interchangeable with the GAITRite mat.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Gait & Posture - Volume 38, Issue 3, July 2013, Pages 380-384
نویسندگان
, , ,