کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
6209012 1603982 2015 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Better Axial Stiffness of a Bicortical Screw Construct Compared to a Cable Construct for Comminuted Vancouver B1 Proximal Femoral Fractures
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Better Axial Stiffness of a Bicortical Screw Construct Compared to a Cable Construct for Comminuted Vancouver B1 Proximal Femoral Fractures
چکیده انگلیسی

The aim of this study was to biomechanically evaluate the Locking attachment plate (LAP) construct in comparison to a Cable plate construct, for the fixation of periprosthetic femoral fractures after cemented total hip arthroplasty. Each construct incorporated a locking compression plate with bi-cortical locking screws for distal fixation. In the Cable construct, 2 cables and 2 uni-cortical locking screws were used for proximal fixation. In the LAP construct, the cables were replaced by a LAP with 4 bi-cortical locking screws. The LAP construct was significantly stiffer than the cable construct under axial load with a bone gap (P = 0.01). The LAP construct offers better axial stiffness compared to the cable construct in the fixation of comminuted Vancouver B1 proximal femoral fractures.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 30, Issue 12, December 2015, Pages 2333-2337
نویسندگان
, , , , , , , , , ,