کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
6209735 1604011 2013 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Tibial Cementing in UKA: A Three-Dimensional Analysis of the Bone Cement Implant Interface and the Effect of Bone Lavage
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Tibial Cementing in UKA: A Three-Dimensional Analysis of the Bone Cement Implant Interface and the Effect of Bone Lavage
چکیده انگلیسی

Loosening is a common cause for revision in cemented UKA. In a cadaver study, we analyzed the three-dimensional cement distribution under the tibial implant and the effect of bone lavage (pulsed lavage, syringe lavage) on maximum cement penetration and penetration volume. Analyses were determined by performing bone cuts in medio-lateral direction and converting this data into a 3D model. Pulsed lavage led to an increased mean maximum cement penetration 5.79 ± 2.63 mm and penetration volume 6471.34 ± 1156.43 mm3 compared to syringe lavage 4.62 ± 2.61 mm, 5069.81 ± 1177.09 mm3 (P < 0.001; P < 0.001). Our results show a complete cement mantle for both investigated lavage techniques. Cleansing the cancellous tibial bone bed using pulsed lavage is more effective than conventional syringe lavage and leads to a deeper cement penetration and lager cement penetration volume under the tibial component.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 28, Issue 9, Supplement, October 2013, Pages 191-194
نویسندگان
, , , , , , ,