Accurate implant positioning and restoration of lower limb alignment are major requirements for successful long-term results in unicompartmental knee arthroplasty (UKA). Alignment accuracy was compared between navigated-UKA (nUKA) and conventional-UKA (cUKA) groups using a retrospective matched case-control study (n = 129, 58 nUKA, 71 cUKA). Mechanical axis (MA), hip-knee-ankle angle (HKA°), coronal implant alignment, and tibial implant posterior slope were measured. No statistically significant difference was observed when comparing MA, HKA° or coronal implant alignment (p > 0.05). Statistical significance was seen with tibial component posterior slope (p = 0.04, nUKA 4.2°, cUKA 2.9°); and between intra-operative navigationally determined HKA° and post-operative whole-leg standing (WLS) film HKA°. Navigation does not significantly improve UKA alignment compared to conventional methods. Further studies are needed to justify the use of this technology in UKA.
Journal: The Journal of Arthroplasty - Volume 28, Issue 3, March 2013, Pages 463-468