کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
6210002 1604018 2013 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Unsatisfactory Accuracy as Determined by Computer Navigation of VISIONAIRE Patient-Specific Instrumentation for Total Knee Arthroplasty
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Unsatisfactory Accuracy as Determined by Computer Navigation of VISIONAIRE Patient-Specific Instrumentation for Total Knee Arthroplasty
چکیده انگلیسی

This investigation evaluated the Smith and Nephew VISIONAIRE patient-specific cutting block (PSCB) system for total knee arthroplasty. A consecutive series of 60 patients was recruited. Intraoperative computer navigation was used to evaluate the accuracy of the cutting blocks in the coronal and sagittal planes for the tibia, as well as rotational plane for the femur. The PSCB would have placed 79.3% of the sample within ± 3° of the preoperative plan in the coronal plane, while the rotational and sagittal alignment results within ± 3° were 77.2% and 54.5% respectively. The VISIONAIRE PSCB system achieved unacceptable accuracy when assessed by computer navigation. There might be many sources of error, but caution is recommended before using this system routinely without objective verification of alignment.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 28, Issue 3, March 2013, Pages 469-473
نویسندگان
, , , , , , ,