کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
6211537 | 1267240 | 2009 | 4 صفحه PDF | دانلود رایگان |

The St. Leger total knee replacement (Zynergy Orthopaedics Ltd, Rotherham, UK) was developed as a cheaper alternative to similar implants of its time. Between October 1993 and June 1999, 144 St. Leger total knee replacements were implanted into 114 patients. Seventy-three patients (99 knees) were recalled for assessment (mean follow-up of 10.2Â years). Eighteen patients had had their prostheses revised, 11 had died and 12 were lost to follow-up. Functional Score showed 90% poor results and the Objective Knee Score showed 31% poor results. Radiological assessment identified 12 arthroplasties that had failed and 58 that required close follow. Kaplan-Meier cumulative survivorship was 87% at 10Â years. The St. Leger knee replacement did not perform as well as others of the same generation and was not worth the initial financial savings.
Journal: The Knee - Volume 16, Issue 5, October 2009, Pages 322-325