کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
711272 | 892127 | 2015 | 7 صفحه PDF | دانلود رایگان |
Diabetes associated complications are affecting an increasingly large population of hospitalized patients. Since glucose physiology is significantly impacted by patient-specific parameters, it is critical to verify that a clinical glucose control protocol is safe across a wide patient population. A safe protocol should not drive the glucose level into dangerous low (hypoglycemia) or high (hyperglycemia) ranges. Verification of glucose controllers is challenging due to the high-dimensional, non-linear glucose physiological models which contain both unobservable states and unmeasurable patient-specific parameters. This paper presents a hybrid system model of a closed-loop physiological system that includes an existing FDA-accepted high-fidelity physiological model tailored to intraoperative settings and a validated improvement to a clinical glucose control protocol for diabetic cardiac surgery patients. We propose the closed-loop model as a physiological system benchmark for verification and present our initial results on verifying the system using the SMT-based hybrid system verification tool dReach.
Journal: IFAC-PapersOnLine - Volume 48, Issue 27, 2015, Pages 211-217