کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
731435 | 893065 | 2012 | 5 صفحه PDF | دانلود رایگان |

This paper describes the procedures made at National Institute for Standards (NIS), Egypt to establish accurate resistance measurements for high value resistance standards in the range from 100 kΩ to 100 MΩ. Guarded Hamon transfer standards have been used at NIS for scaling to high resistance levels up to 100 MΩ. The used method at NIS for the high resistance measurements, above 100 kΩ, is the DMM-based method which works automatically. A full description of the used method and of the new software which is made especially for this purpose are reported. Also, the uncertainty for the measurement of the Hamon transfer standard used for the scaling process and the standard resistor under calibration are evaluated.
► Establish high valued resistance measurement at NIS from 100 kΩ to 100 MΩ.
► NIS 10 kΩ and two Hamon standards are used for the scaling process.
► Automated DMM-based method are used.
► Special program is done for this purpose.
Journal: Measurement - Volume 45, Issue 5, June 2012, Pages 988–992