کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
8167503 | 1526249 | 2018 | 17 صفحه PDF | دانلود رایگان |
عنوان انگلیسی مقاله ISI
A gated LaBr3(Ce) detector for border protection applications
دانلود مقاله + سفارش ترجمه
دانلود مقاله ISI انگلیسی
رایگان برای ایرانیان
موضوعات مرتبط
مهندسی و علوم پایه
فیزیک و نجوم
ابزار دقیق
پیش نمایش صفحه اول مقاله

چکیده انگلیسی
We report on the dedicated implementation of the blocking technique for a LaBr3(Ce) detector as well as associated electronics and data acquisition system for border protection applications. The detector is meant to perform delayed γ-ray spectroscopy of fission fragments produced via photofission induced by a high intensity pulsed photon beam. The gating technique avoids saturation of the detection chain during irradiation. The resulting setup allows us to successfully perform delayed γ-ray spectroscopy starting only 30 ns after the gating operation. The measured energy resolution ranges from 5% to 6.5% at 662  keV depending on the γ-ray detection time after the gating operation.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 877, 1 January 2018, Pages 323-327
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 877, 1 January 2018, Pages 323-327
نویسندگان
A. Etile, D. Denis-Petit, L. Gaudefroy, V. Meot, O. Roig,