Investigating the lifetime of bromine-quenched G.M. Counters with temperature
ترجمه فارسی عنوان
بررسی طول عمر باکتری برم شمارنده با درجه حرارت
کلمات کلیدی
گیگر گاز پر شده، هالوژن کاشی کاری، طول عمر، درجه حرارت،
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
چکیده انگلیسی
The amount of halogen quench gas as a percentage of the total fill gas contained within a gas-filled Geiger-Müller detector is directly linked to its operational characteristics. Preserving this halogen gas will help maintain the operating lifetime of the detectors. Such halogen gases are highly corrosive and the small quantities within a gas-filled detector can deplete rapidly via interactions with surrounding materials. The rate of interactions is thought to be proportional to not only temperature, but also to the current initiated by ionisation events associated with the formation of each signal pulse. As all pulses are of similar magnitudes, GM detector operational lifetimes are quantified in accumulated counts rather than a given operating time. We have studied three different types of corrosion-resistant mechanisms to protect the bromine halogen gas from any interactions with 446 stainless steel detector components of ZP1200 Geiger-Müller tubes at temperatures up to 125 °C. Three types of surface treatments for these detectors used were labelled as “raw” using only an oxygen-plasma-bombardment process, “passivated” using a combination of nitric acid passivation followed by an oxygen plasma-bombardment process, and thirdly plating with a few micron thickness of chromium followed by an oxygen plasma-bombardment process. 32 detector samples were irradiated at room temperature with a Caesium-137 source at dose rates of approximately 1.3 mSv/hr up to 5.7×1010 accumulated counts; another 32 detector samples were aged to 3.3×1010 counts at 125 °C. At room temperature, the chromium-plated samples exhibited an initial rise in their starting voltage readings. All other detector performance characteristics, for all detector types, did not change significantly during the ageing process, and the surface morphology of the detector cathodes was unaffected. At 125 °C, the chromium-based plating produced the most stable long-term response. These chromium-plated samples showed no evidence of bromine deposits on the cathode samples investigated via Energy-Dispersive X-ray Spectroscopy. The raw and passivated samples, however, did show traces of bromine independent of the age of the detector. Preliminary investigations we have carried out at 175 °C highlight the importance of the chromium plating layer in preserving long-term detector performances at elevated temperatures.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 795, 21 September 2015, Pages 12-18
نویسندگان
, , ,