کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8208562 1532060 2018 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
QA measurement of gamma-ray dose and neutron activation using TLD-400 for BNCT beam
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
QA measurement of gamma-ray dose and neutron activation using TLD-400 for BNCT beam
چکیده انگلیسی
TLD-400 (CaF2:Mn) chips were applied for the gamma-ray dose measurement in a PMMA phantom exposed to a BNCT beam because of their very low neutron sensitivity. Since TLD-400 chips possess an adequate amount of Mn activator they have been employed in this work simultaneously for neuron activation measurement. The self-irradiation TL signals owing to the decay of the neutron induced 56Mn activity have been applied for a calibration of the TLD-400 chip in situ, where the activities were measured by an HPGe detector system and the energy deposition per disintegration of 56Mn was calculated by applying a Monte Carlo code. It was accidentally found that the irradiated TLD-400 chips were capable of emitting prominent scintillation lights owing to the induced 56Mn activity, which can easily be recorded by the TLD reader without heating and after a calibration can be used to determine the 56Mn activity.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 137, July 2018, Pages 73-79
نویسندگان
, , ,