کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8208749 1532064 2018 25 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Application of the 'Spiking' method to the measurement of low dose radiation (≤ 1 Gy) using alanine dosimeters
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Application of the 'Spiking' method to the measurement of low dose radiation (≤ 1 Gy) using alanine dosimeters
چکیده انگلیسی
Alanine dosimeters are limited in radiotherapy by poor sensitivity at low doses (< 5 Gy). A set of alanine dosimeters were 'spiked' with a large dose of radiation, (~30 Gy, 6 MV X-rays) and additional doses ranging between 0.5 and 10 Gy. The radical yield was measured by Electron Paramagnetic Resonance (EPR) spectroscopy, and after subtraction of the contribution from the "spike" dose, a linear correlation between the radiation dose and the area of the central EPR signal was obtained for doses between 0.5 and 10 Gy (regression value of 0.9890), and for the central peak's amplitude (regression value of 0.9895). Overall, this method is easy to perform, requires no complex EPR signal analysis, and offers much potential to extend the current usage of alanine dosimeters in radiotherapy.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 133, March 2018, Pages 111-116
نویسندگان
, , , , , ,