Calibration and efficiency curve of SANAEM ionization chamber for activity measurements
چکیده انگلیسی
A commercially available Fidelis ionization chamber was calibrated and assessed in PTB with activity standard solutions. The long-term stability and linearity of the system was checked. Energy-dependent efficiency curves for photons and beta particles were determined, using an iterative method in Excel™, to enable calibration factors to be calculated for radionuclides which were not used in the calibration. Relative deviations between experimental and calculated radionuclide efficiencies are of the order of 1% for most photon emitters and below 5% for pure beta emitters. The system will enable TAEK-SANAEM to provide traceable activity measurements.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 109, March 2016, Pages 70-73