کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8209561 1532091 2015 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
PGNAA system preliminary design and measurement of In-Hospital Neutron Irradiator for boron concentration measurement
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
PGNAA system preliminary design and measurement of In-Hospital Neutron Irradiator for boron concentration measurement
چکیده انگلیسی
A prompt gamma neutron activation analysis (PGNAA) system has been recently developed at the 30-kW research reactor In-Hospital Neutron Irradiator (IHNI) in Beijing. Neutrons from the specially designed thermal neutron beam were used. The thermal flux of this beam is 3.08×106 cm−2 s−1 at a full reactor power of 30 kW. The PGNAA system consists of an n-type high-purity germanium (HPGe) detector of 40% efficiency, a digital spectrometer, and a shielding part. For both the detector shielding part and the neutron beam shielding part, the inner layer is composed of 6Li2CO3 powder and the outer layer lead. The boron-10 sensitivity of the PGNAA system is approximately 2.5 cps/ppm. Two calibration curves were produced for the 1-10 ppm and 10-50 ppm samples. The measurement results of the control samples were in accordance with the inductively coupled plasma atomic emission spectroscopy (ICP-AES) results.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 106, December 2015, Pages 161-165
نویسندگان
, , , , , ,