Low temperature gamma sterilization of a bioresorbable polymer, PLGA
چکیده انگلیسی
The results showed that the molecular weight was significantly reduced as was the glass transition temperature, an indication of chain scission. FTIR showed small changes in chemical structure in the methyl and carbonyl groups after irradiation. Glass transition temperature was significantly different between irradiation at −80 °C and irradiation at 25 °C, however this was a difference of only 1 °C. Ultimately, the results indicate that the sterilization temperature used does not affect PLGA when carried out in a nitrogen atmosphere.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiation Physics and Chemistry - Volume 143, February 2018, Pages 27-32