کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
8252520 | 1533508 | 2015 | 5 صفحه PDF | دانلود رایگان |
عنوان انگلیسی مقاله ISI
A new G-M counter hybrid dead-time correction model
دانلود مقاله + سفارش ترجمه
دانلود مقاله ISI انگلیسی
رایگان برای ایرانیان
کلمات کلیدی
موضوعات مرتبط
مهندسی و علوم پایه
فیزیک و نجوم
تشعشع
پیش نمایش صفحه اول مقاله

چکیده انگلیسی
A new G-M counter dead-time model based on the hybrid model is proposed and studied. The new model contains the idealized paralyzable and non-paralyzable dead-time models, but also allows the non-paralyzable dead-time to vary with the true counting rate. To validate this model, the decaying source experiment with a very pure 56Mn source was used. The calculated true counting rates obtained by fitting the experimental data to the model were compared to the data obtained by table lookup. In general, the fitting was in good agreement (within 4.8% up to 5.35Ã106counts/minute) for a G-M counter with a paralyzable dead-time component of 85.9 μs and a non-paralyzable dead-time component ranging from 438 μs to 24.3 ms.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiation Physics and Chemistry - Volume 116, November 2015, Pages 125-129
Journal: Radiation Physics and Chemistry - Volume 116, November 2015, Pages 125-129
نویسندگان
Guojing Hou, Robin P. Gardner,