On timing response improvement of an NE213 scintillator attached to two PMTs
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
On timing response improvement of an NE213 scintillator attached to two PMTs
چکیده انگلیسی
A 5 cm diameter by 6 cm height NE213 scintillator attached to two XP2282 PHOTONIS photomultiplier tubes (PMTs) exposed to 241Americium-Berylium (Am-Be) neutron-gamma source has been used for timing response studies. The neutron-gamma discrimination (NGD) measurements based on a modified zero-crossing (ZC) method show that the discrimination quality, usually expressed in figure-of-merit (FoM) and peak-to-valley (P/V) values, has been improved. The timing response evaluated with Monte Carlo light transport code, PHOTRACK, also verifies this improvement.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiation Physics and Chemistry - Volume 90, September 2013, Pages 6-10