کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8737342 1591331 2018 23 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Comparison of Abbott Architect®, Siemens Immulite®, and Diasorin Liaison® for determination of Epstein-Barr virus serological diagnosis
موضوعات مرتبط
علوم زیستی و بیوفناوری ایمنی شناسی و میکروب شناسی میکروبیولوژی و بیوتکنولوژی کاربردی
پیش نمایش صفحه اول مقاله
Comparison of Abbott Architect®, Siemens Immulite®, and Diasorin Liaison® for determination of Epstein-Barr virus serological diagnosis
چکیده انگلیسی
Although similar results were obtained with Architect® and Liaison®, several discordant results were observed with Immulite®, particularly for EBNA IgG. A large number of EBNA IgG-positive results were observed, which interfered with interpretation of the EBV profile. In contrast, Immulite® performed similarly to the 2 other assays for detection of VCA IgM.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Diagnostic Microbiology and Infectious Disease - Volume 90, Issue 2, February 2018, Pages 96-101
نویسندگان
, , , , , , , , , , ,