کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
874402 910336 2007 10 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Experimental validation of intact and implanted distal femur finite element models
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی پزشکی
پیش نمایش صفحه اول مقاله
Experimental validation of intact and implanted distal femur finite element models
چکیده انگلیسی

Four finite element (FE) models of intact and distal femur of knee replacements were validated relative to measured bone strains. FE models of linear tetrahedrons were used. Femoral replacements with cemented stemless, cemented and noncemented femoral stems of the PFC Sigma Modular Knee System were analyzed. Bone strains were recorded at ten locations on the cortex. The magnitude of the FE bone strains corresponded to the mean measured strains, with an overall agreement of 10%. Linear regression between the FE and mean experimental strains produced slopes between 0.94 and 1.06 and R2 values between 0.92 and 0.99. RSME values were less than 12%. The FE models were able to adequately replicate the mechanical behavior of distal femur reconstructions.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Biomechanics - Volume 40, Issue 11, 2007, Pages 2467–2476