کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8932149 1644551 2015 9 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Comparison of pulse contour analysis by Pulsioflex and Vigileo to measure and track changes of cardiac output in critically ill patients
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی بیهوشی و پزشکی درد
پیش نمایش صفحه اول مقاله
Comparison of pulse contour analysis by Pulsioflex and Vigileo to measure and track changes of cardiac output in critically ill patients
چکیده انگلیسی
The Pulsioflex did not reliably estimate the absolute values of CI. For tracking fluid-induced changes in CI, the Pulsioflex was reliable, and also the Vigileo. For tracking norepinephrine-induced changes in CI, it was also reliable and tended to be better than the Vigileo. Auto-calibration allowed by the system did not improve its reliability.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: British Journal of Anaesthesia - Volume 114, Issue 2, February 2015, Pages 235-243