Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurement
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی بیهوشی و پزشکی درد
پیش نمایش صفحه اول مقاله
Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurement
چکیده انگلیسی
The CNAP monitor showed an acceptable agreement and was interchangeable with invasive pressure monitoring for MAP during normotensive conditions. During induction of anaesthesia and when the AP was low, the agreement was less good and interchangeability was not achieved. These results suggest that CNAP is not statistically equivalent to invasive monitoring during all periods of anaesthesia but may be a useful additional AP monitor.