کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8935921 1644604 2011 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Comparison of two methods for cardiac output measurement in critically ill patients†
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی بیهوشی و پزشکی درد
پیش نمایش صفحه اول مقاله
Comparison of two methods for cardiac output measurement in critically ill patients†
چکیده انگلیسی
Our data show a poor level of concordance between measures provided by the two methods. We found an underestimation of CO values measured with the 1.07 software version of FloTrac for supranormal values of CO. The new software (1.10) has been improved in order to correct this bias; however, its reliability is still poor. On the basis of our data, we can therefore conclude that both software versions of FloTrac/Vigileo did not still provide reliable estimation of CO in our intensive care unit setting.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: British Journal of Anaesthesia - Volume 106, Issue 5, May 2011, Pages 690-694
نویسندگان
, , , , , , ,