Evaluation of a one-step sample preparation protocol for analysis of total serum testosterone by LC-MS/MS
دانلود مقاله + سفارش ترجمه
دانلود مقاله ISI انگلیسی
رایگان برای ایرانیان
کلمات کلیدی
LLOQTESTODHEANH4OACEtOAcMPBdehydroepiandrosteroneMTBEMPAMRMDHT1-Chlorobutane - 1-کلروبوتان4-dione - 4-دیئون5-diol - 5-دیولLC–MS/MS - LC-MS / MSandrost-5-ene-3β,17β-diol - آندروست-5-آن-3β، 17 بتا دیولEthyl acetate - اتیل استاتAmmonium acetate - استات آمونیومLiquid-liquid extraction - استخراج مایع _ مایعandrostenedione - اندرسوندیونtestosterone - تستوسترونDihydrotestosterone - دی هیدروتستوسترونmethyl tert-Butyl ether - متیل ترتل-بوتیل اترlower limit of quantitation - محدودیت پایین کمیتmultiple reaction monitoring - نظارت چندگانه چندگانهliquid chromatography–tandem mass spectrometry - کروماتوگرافی مایع و اسپکترومتری توده دو طرفهquality control - کنترل کیفیت
موضوعات مرتبط
مهندسی و علوم پایه
شیمی
طیف سنجی
پیش نمایش صفحه اول مقاله

چکیده انگلیسی
In this study, a one-step method for liquid-liquid extraction has been compared against a two-step procedure for testosterone assays in terms of accuracy, specificity, recovery, lipid removal and baseline noise, using QCs and unknown samples. The difference in accuracy was less than 5% for adult sera, while it was less than 10% for prepubescent sera. To compare specificity, the ion ratio transition of 289â¯ââ¯97 to 289â¯ââ¯109 was monitored for all QCs and unknown samples; no interference in the testosterone peak was observed for any tested sample prepared by either the one-step or two-step procedure. The baseline comparison of LC-MS/MS chromatograms of samples indicated that samples prepared by the one-step procedure were of the same quality as those prepared by the two-step procedure; however, recovery in unaltered serum using the one-step procedure was approximately 15% greater across the low to high concentration range. Furthermore, recovery using the one-step procedure was more consistent between stripped and unstripped serum. Lipids were removed efficiently in the two-step procedure as verified by monitoring the typical phospholipid MRM transition of m/z 496â¯ââ¯184. For every sample processed by the one-step procedure, a one-minute online column wash with 95% methanol was able to remove 95% of the bound lipids, thereby providing a column life-time approximately equivalent to that for the two-step procedure. The presented data indicate that the one-step procedure could replace the two-step procedure while maintaining accuracy, saving time, increasing recovery, and minimizing the potential for errors with the fewer steps required.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Clinical Mass Spectrometry - Volume 10, December 2018, Pages 9-15
Journal: Clinical Mass Spectrometry - Volume 10, December 2018, Pages 9-15
نویسندگان
Yuyong Ke, Alain Dury, Renaud Gonthier, Fernand Labrie,