A comparison of CAT Doppler and oscillometric Memoprint machines for non-invasive blood pressure measurement in conscious cats
دانلود مقاله + سفارش ترجمه
دانلود مقاله ISI انگلیسی
رایگان برای ایرانیان
موضوعات مرتبط
علوم زیستی و بیوفناوری
علوم کشاورزی و بیولوژیک
علوم دامی و جانورشناسی
پیش نمایش صفحه اول مقاله

چکیده انگلیسی
Indirect blood pressure measurements were compared in 28 conscious cats using Doppler and oscillometric blood pressure-measuring devices. Ten cats were used to compare Doppler measurements between two examiners and 18 cats were used to compare Doppler and oscillometric measurements. The Doppler machine obtained systolic and diastolic blood pressure readings in 100% and 51% of attempts, respectively. With the oscillometric machine, systolic and diastolic blood pressure readings were obtained in 52% of the attempts. With the Doppler, measures of mean systolic blood pressure between two examiners were positively correlated, but there was no correlation for diastolic blood pressure measures. When comparing the results obtained by Doppler and oscillometric machines there was no significant difference between mean systolic blood pressure readings, but the oscillometric machine produced significantly higher estimates of diastolic blood pressure. In both cases, the standard deviations for the oscillometric machine were considerably larger than those for the Doppler machine. The first reading of systolic blood pressure obtained with the Doppler machine was an excellent predictor of the mean of five readings, but this was not so for the oscillometric machine. It took less than 5Â min to obtain five readings in 37.5% of cases with the Doppler machine but this was true for only 5% of cases with the oscillometric machine. Two cats with ophthalmological lesions consistent with systemic hypertension were identified. In these two patients, systolic blood pressure measurements were between 200 and 225Â mmHg when measured by Doppler, and between 140 and 150Â mmHg when measured by the oscillometric machine. This suggests that a lower reference range for normal systolic blood pressure values should be used for the oscillometric device.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Feline Medicine & Surgery - Volume 7, Issue 3, June 2005, Pages 147-152
Journal: Journal of Feline Medicine & Surgery - Volume 7, Issue 3, June 2005, Pages 147-152
نویسندگان
Rosanne E. BVSc, MRCVS, Vivien BSc, BVSc, MRCVS, Michael BA (Hons), PhD, Sarah M.E. BVSc, PhD, DSAM (Feline), MRCVS, David J. BSc (Hons), BVM&S, PhD, DVOphthal, DipECVO, MRCVS,