None of the 14 out of 16 successfully tested legs or screws failed. No difference was found in fixation of the syndesmosis when stainless steel screws were compared to titanium screws through three or four cortices. Mean lateral displacement found after testing was 1.05Â mm (S.D. = 0.42). This increase in tibiofibular width exceeds values described in literature for the intact syndesmosis loaded with body weight. Based on this laboratory study it is concluded that the syndesmotic set screw cannot prevent excessive syndesmotic widening when loaded with a load comparable with body weight. Therefore, we advise that patients with a syndesmotic set screw in situ should not bear weight.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Injury - Volume 36, Issue 1, January 2005, Pages 60-64
Journal: Injury - Volume 36, Issue 1, January 2005, Pages 60-64
نویسندگان
Annechien Beumer, Martin M. Campo, Ruud Niesing, Judd Day, Gert-Jan Kleinrensink, Bart A. Swierstra,