کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
97432 | 160493 | 2008 | 5 صفحه PDF | دانلود رایگان |

Headspace solid-phase microextraction (HS-SPME) is assessed as an alternative to liquid–liquid extraction (LLE) currently used for 3,4-methylenedioxymethampethamine (MDMA) profiling. Both methods were compared evaluating their performance in discriminating and classifying samples. For this purpose 62 different seizures were analysed using both extraction techniques followed by gas chromatography–mass spectroscopy (GC–MS). A previously validated method provided data for HS-SPME, whereas LLE data were collected applying a harmonised methodology developed and used in the European project CHAMP.After suitable pre-treatment, similarities between sample pairs were studied using the Pearson correlation.Both methods enable to distinguish between samples coming from the same pre-tabletting batches and samples coming from different pre-tabletting batches. This finding emphasizes the use of HS-SPME as an effective alternative to LLE, with additional advantages such as sample preparation and a solvent-free process.
Journal: Forensic Science International - Volume 182, Issues 1–3, 20 November 2008, Pages 52–56