کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10429603 909845 2005 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Amperometric microbial biosensor for p-nitrophenol using Moraxella sp.-modified carbon paste electrode
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Amperometric microbial biosensor for p-nitrophenol using Moraxella sp.-modified carbon paste electrode
چکیده انگلیسی
An amperometric microbial biosensor for highly specific, sensitive and rapid quantitative determination of p-nitrophenol was developed. The biosensor takes advantage of the ability of Moraxella sp. to specifically degrade p-nitrophenol to hydroquinone, a more electroactive compound than p-nitrophenol. The electrochemical oxidation current of hydroquinone formed in biodegradation of p-nitrophenol was measured at Moraxella sp.-modified carbon paste electrode and correlated to p-phenol concentrations. The optimum response was realized by electrode constructed using 15 mg of dry cell weight per 1 g of carbon paste and operating at 0.3 V (versus Ag/AgCl reference) in pH 7.5, 20 mM sodium phosphate buffer. Operating at these optimum conditions the biosensor had excellent selectivity against phenol derivatives and was able to measure as low as 20 nM (2.78 ppb) p-nitrophenol with very good accuracy and reproducibility. The biosensor was stable for approximately 3 weeks when stored at 4 °C. The applicability of the biosensor to measure p-nitrophenol in lake water was demonstrated.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Biosensors and Bioelectronics - Volume 21, Issue 3, 15 September 2005, Pages 523-527
نویسندگان
, , , , ,