کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10429879 909868 2005 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
QCM-FIA with PGMA coating for dynamic interaction study of heparin and antithrombin III
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
QCM-FIA with PGMA coating for dynamic interaction study of heparin and antithrombin III
چکیده انگلیسی
In this work, we describe a method of constructing a film of linear poly(glycidyl methacrylate) (PGMA) polymer onto the surface of quartz crystal microbalance (QCM) electrode as a coating material that allows easy coupling of heparin molecules onto the electrode and facilitates the determination of the interaction between heparin and antithrombin III (AT III). The PGMA film was characterized with atomic force microscopy (AFM) and infra-red spectroscopy. The coupling of heparin was accomplished in one step solution reaction. A home-made quartz crystal microbalance-flow injection analysis (QCM-FIA) system with data analysis software developed in our laboratory was used to determine the interaction. The interactions between immobilized heparin and AT III were studied with various concentrations under various conditions. The obtained constants are kass = (1.49 ± 0.12) × 103 mol−1 l s−1, kdiss = (3.94 ± 0.63) × 10−2 s−1, KA = (3.82 ± 0.33) × 104 mol−1 l.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Biosensors and Bioelectronics - Volume 21, Issue 1, 15 July 2005, Pages 121-127