Influence of clearance on the time-dependent performance of the hip following hemiarthroplasty: A finite element study with biphasic acetabular cartilage properties
ترجمه فارسی عنوان
تأثیر ترخیص بر عملکرد وابسته به زمان همیادرفروپلاستی مفصل ران: یک مطالعه عنصر محدود با خواص غضروف استاپول دو طرفه
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی پزشکی
چکیده انگلیسی
Hip hemiarthroplasty is a common treatment for femoral neck fracture. However, the acetabular cartilage may degenerate after hemiarthroplasty leading to postoperative failure and the need for revision surgery. The clearance between the acetabular cartilage and head of the prosthesis is one of the potential reasons for this failure. In this study, the influence of joint clearance on the biomechanical function of a generic hip model in hemiarthroplasty was investigated using biphasic numerical simulation. Both a prolonged loading period of 4000 s and dynamic gait load of 10 cycles were considered. It was found that a larger clearance led to a higher stress level, a faster reduction in load supported by the fluid and a faster cartilage consolidation process. Additionally, the mechanical performance of the acetabular cartilage in the natural model was similar to that in the hemiarthroplasty model with no clearance but different from the hemiarthroplasty models with clearances of 0.5 mm and larger. The results demonstrated that a larger clearance in hip hemiarthroplasty is more harmful to the acetabular cartilage and prosthesis heads with more available dimensions (i.e. smaller increments in diameter) could be manufactured for surgeons to achieve a lower clearance, and reduced contact stress in hemiarthroplasty surgeries.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Medical Engineering & Physics - Volume 36, Issue 11, November 2014, Pages 1449-1454
نویسندگان
, , , , ,