Quantitation of aldosterone in human plasma by ultra high performance liquid chromatography tandem mass spectrometry
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Quantitation of aldosterone in human plasma by ultra high performance liquid chromatography tandem mass spectrometry
چکیده انگلیسی

Aldosterone is a mineralocorticoid steroid hormone whose measurement in the clinical laboratory is principally performed for the investigation of primary hyperaldosteronism. Traditionally measurement of aldosterone has been performed by radioimmunoassay, however these assays lack specificity as they are prone to interference from structurally related steroid hormones. Herein, we have developed a novel, sensitive and specific method utilising solid phase extraction and quantitation of aldosterone from human plasma by UPLC–MS/MS. Standards, quality controls and samples (250 μL) were extracted using Oasis® HLB 96-well plates. Extract (30 μL) was injected onto a Krudcatcher UPLC In-Line Filter, 0.5 μm guard column, coupled to a Kinetex PFP, 100 mm × 2.1 mm, 2.6 μm column with methanolic mobile phase gradient elution. Eluant was connected to a Waters® Xevo TQS tandem mass spectrometer operating in electrospray negative mode. We detected multiple reaction monitoring (MRM) transitions of m/z 359.0 > 189.1 for aldosterone and 366.0 > 194.1 for d7-aldosterone respectively, which co-eluted at 2.65 min. Ion suppression was negligible. Mean recovery was 89.6%, limit of detection and lower limit of quantitation were 26 pmol/L and 30 pmol/L respectively. The assay was linear up to 3200 pmol/L (r2 = 0.9999). Mean intra- and inter-assay imprecision and bias were all <10%. Comparison of the UPLC–MS/MS method with an immunoassay in routine clinical use in the UK yielded the equation UPLC–MS/MS = 0.789(RIA)–41.7, linear regression r2 = 0.88, n = 54. We have developed a sensitive and specific method for the extraction and measurement of aldosterone from human plasma. The method features a simple 96-well plate solid phase extraction procedure, highly selective column chemistry and short chromatographic run times.


► We report a novel UPLC–MS/MS method for the quantitation of aldosterone from human plasma.
► Sample extraction utilises a rapid solid phase extraction technique.
► Chromatography uses highly selective pentaflurophenyl functional groups.
► Short chromatographic run times enable high throughput analysis of clinical samples.
► The LC–MS/MS method demonstrates improved specificity compared to immunoassay.