کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1213236 | 1494137 | 2012 | 6 صفحه PDF | دانلود رایگان |

In this work we compared the performance of chromatography columns with particles of 5 and 3 μm with the new 2.7 μm solid core particles for the analysis of aflatoxins B1, G1, B2, and G2 using trifluoroacetic acid pre-column derivatization. Three different columns have been used and chromatographic parameters as retention time, resolution, limit of detection (LOD), limit of quantification (LOQ) were obtained from all of them and compared. The results show that comparing with the traditional columns, shorter columns (100 mm × 4.6 mm) with the new solid core particles are suitable for the analysis of these mycotoxins and allowed the reduction of the analysis time by 45.5% and 33.3% with respect to columns with particle size 5 μm (150 mm × 4.6 mm) and 3 μm (150 mm × 4.6 mm) respectively, without any detrimental effect on performance. This leads to the reduction of the analysis costs by saving on organic solvents and increasing the total number of analyses per day. The capability of these columns for analyzing samples, in different culture media, was assessed by analyzing different samples from: yeasts extract sucrose medium, corn meal agar medium and fresh hazelnut media.
► Different C18 particle sizes have been compared to analyse aflatoxins.
► Solid core particles allowed the reduction of the analysis time.
► Reductions of analysis costs were achieved by reducing time and solvents use.
► The suitability to analyse aflatoxin from different matrixes was demonstrated.
Journal: Journal of Chromatography B - Volumes 889–890, 15 March 2012, Pages 138–143