کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1213666 | 966894 | 2010 | 6 صفحه PDF | دانلود رایگان |

A sensitive and reproducible HPLC method for quantitative determination of puerarin (PUE) in rat plasma was developed and validated using 4-hydroxybenzaldehyde as an internal standard. The separation of PUE was performed on a CAPCELL PAK C18 column by gradient elution with 0.2% aqueous phosphoric acid and acetonitrile as the mobile phase. The method was validated and found to be linear in the range of 80–12,000 ng/mL. The limit of quantification was 80 ng/mL based on 100 μL of plasma. The variations for intra- and inter-day precision were less than 8.3%, and the accuracy values were between 98% and 105.2%. The extraction recoveries were more than 85%. The method was successfully applied in the comparative study of pharmacokinetics of PEGylated puerarin (PEG-PUE) versus PUE in rats. Compared with PUE, PEG-PUE showed a 5.2-fold increase in half-life of PUE and a 4.7-fold increase in mean residence time. In addition, this method was also successfully applied to determine the low plasma concentration of PUE regenerated from PEG-PUE in vitro.
Journal: Journal of Chromatography B - Volume 878, Issue 31, 1 December 2010, Pages 3297–3302