کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1214101 | 966918 | 2010 | 6 صفحه PDF | دانلود رایگان |
A sensitive and high throughput off-line μElution 96-well solid-phase extraction (SPE) followed by strong cation exchange (SCX) liquid chromatography with tandem mass spectrometry (LC/MS/MS) quantification for determination of cefepime has been developed and validated in mouse plasma. Using the chemical analog, ceftazidime as an internal standard (IS), the linear range of the method for the determination of cefepime in mouse plasma was 4–2048 ng/mL with the lower limit of quantitation level (LLOQ) of 4 ng/mL. The inter- and intra-assay precision and accuracy of the method were below 9.05% and ranged from 95.6 to 113%, respectively, determined by quality control (QC) samples at five concentration levels including LLOQ. After μElution SPE, 71.1% of cefepime was recovered. The application of the validated assay for the determination of cefepime in mouse pharmacokinetics (PK) samples after intravenous (IV) and subcutaneous (SC) doses was demonstrated.
Journal: Journal of Chromatography B - Volume 878, Issue 19, 1 June 2010, Pages 1623–1628