کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1214497 966935 2010 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
2-Propanol in the mobile phase reduces the time of analysis of CLA isomers by silver ion-HPLC
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
2-Propanol in the mobile phase reduces the time of analysis of CLA isomers by silver ion-HPLC
چکیده انگلیسی

Individual isomers of octadecadienoic acid (C18:2) with conjugated double bonds (conjugated linoleic acids; CLA) exert different biological activities. Their distribution in food and tissues differs. Therefore, the separation of the various positional and geometric isomers is important. The time of analysis using silver ion-high performance liquid chromatography can extend up to 90 min. The aim of this study was to reduce this time. The time of analysis reduced from ca. 90 min onto 45 to 35 min, respectively, by the addition of 0.05% or 0.1% (v/v) 2-propanol to the mobile phase [acetonitrile (0.1%; v/v) and diethyl ether (0.5%; v/v) in n-hexane]. There was no effect on resolution of the 17 individual CLA isomers of the CLA mixture. Regarding the lowest coefficient of variation and an adequate baseline separation the use of 0.05% 2-propanol in the mobile phase is recommended, without any disadvantages and adverse effects on the service life of columns. In conclusion, adding 0.05% or 0.1% 2-propanol to the mobile phase shortens the time of analysis of CLA isomers, saves solvents and reduces costs.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography B - Volume 878, Issue 1, 1 January 2010, Pages 88–91
نویسندگان
, , ,