کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1216660 1494192 2006 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Shortcomings of protein removal prior to high performance liquid chromatographic analysis—A case study using method development for BAY 11-7082
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Shortcomings of protein removal prior to high performance liquid chromatographic analysis—A case study using method development for BAY 11-7082
چکیده انگلیسی

During the analytical method development for BAY 11-7082 ((E)-3-[4-methylphenylsulfonyl]-2-propenenitrile), using HPLC-MS–MS and HPLC-UV, we observed that the protein removal process (both ultrafiltration and precipitation method using organic solvents) prior to HPLC brought about a significant reduction in the concentration of this compound. The use of a structurally similar internal standard, BAY 11-7085 ((E)-3-[4-t-butylphenylsulfonyl]-2-propenenitrile), was not effective in compensating for the loss of analyte as the extent of reduction was different to that of the analyte. We present here a systematic investigation of this problem and a new validated method for the determination of BAY 11-7082.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography B - Volume 834, Issues 1–2, 13 April 2006, Pages 93–97
نویسندگان
, , , ,