کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1216808 1494127 2012 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
High-performance liquid chromatography/tandem mass spectrometry for simultaneous determination of four dicaffeoylquinic acids in rat plasma
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
High-performance liquid chromatography/tandem mass spectrometry for simultaneous determination of four dicaffeoylquinic acids in rat plasma
چکیده انگلیسی

A specific and reliable HPLC–MS/MS method was developed and validated for the simultaneous determination of four dicaffeoylquinic acids (DCQA): 3,4-DCQA, 1,5-DCQA, 3,5-DCQA and 4,5-DCQA. The analytes were separated on a C18 column (150 mm × 2.1 mm, 1.8 μm) and a triple-quadrupole mass spectrometry equipped with an electrospray ionization (ESI) source was applied for detection. The plasma sample was prepared by a liquid–liquid extraction method and the recovery for the four analytes was around 80%. The calibration curves were linear over a concentration range of 10.6–1060.0 ng/mL for 3,4-DCQA, 19.2–1920.0 ng/mL for 1,5-DCQA, 14.0–2900.0 ng/mL for 3,5-DCQA, 9.7–970.0 ng/mL for 4,5-DCQA. The intra-day and inter-day precision was less than 15% and the relative error (RE) were all within ±15%. The validated method was successfully applied to a pharmacokinetics study in rats after oral administration of the extracts of Ainsliaea fragrans cham (a traditional Chinese herb).


► Four dicaffeoylquinic acids were the major components of Ainsliaea fragrans cham.
► A C18 column (150 mm × 2.1 mm, 1.8 μm) was applied to separate the four analytes.
► A specific HPLC–MS/MS method was developed to support the pharmacokinetics study.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography B - Volume 903, 15 August 2012, Pages 177–181